  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 087 02 17 000
  • Event Initiated Date
    2017-02-23
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    Safety feature checking the display function may detect a failure leading to a controlled reboot of the cockpit infinity c500 of evita infinity v500 or babylog vn500. as the cockpit infinity c500 operates independently from the ventilation unit evita infinity v500 or babylog vn500, ventilation will continue as set. should the reboot of the cockpit c500 fail, a message “boot disk error” will remain on the screen.

Device

  • Model / Serial
    Cockpit Infinity C500 (MS18746) used for the devices Dräger Infinity ACS Critical Care Workstation (Evita Infinity V500) and Dräger Infinity ACS Neonatal Care Workstation (Babylog VN500).
  • Product Description
    Critical Care Workstation
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturers representative
    Draeger Arabia Co. Ltd.