According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by Philips Healthcare.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 102 02 14 001
  • Event Initiated Date
    2014-05-15
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR, MHRA
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    When the heartstart mrx with q-cpr meter is used. if the user continues to deliver chest compressions when the shock advisory analysis begins or when the device is charging to the selected energy, the ‘do not touch the patient’ icon incorrectly remains on the screen.

Device

  • Model / Serial
    Product: Philips HeartStart MRx Monitor/Defibrillator Model: M3535A and M3536A with Q-CPROM Meter Option B08 Units Affected: MRx units with the Q-CPR Meter option B08 and with a serial number within the range: US00100253 and U500571587. .................... MHRA update code Software revision F.xx, R.xx, T.00.00 or T.00.01 installed.
  • Product Description
    monitor,defibrillators, Q-CPR meters,