translation missing: en.activerecord.attributes.event.titles.recall_field_safety_notice

According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by Philips Healthcare.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 073 10 15 000
  • Event Initiated Date
    2015-10-21
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    Mrx model m3535a with software version f.03.06 and earlier, and model m3536a with version t.00.05 and earlier may stop the automated ready-for-use (rfu) test in an abnormal state when the device is turned off.

Device

  • Model / Serial
    Units Affected: MRx unit types with software versions F.03.06 (M3535A) and T.00.05 (M3536A) and earlier and the following part numbers: PMS Number: 861288 Part Number: M3535A PMS Number: 989803132391 Part Number: M3535ATZ PMS Number: 861289 Part Number: M3536A PMS Number: 989803132401 Part Number: M3536ATZ PMS Number: 861314 Part Number: M3536J PMS Number: 861464 Part Number: M3536M PMS Number: 861481 Part Number: M3536M2 PMS Number: 861483 Part Number: M3536M4 PMS Number: 861484 Part Number: M3536M5 PMS Number: 861491 Part Number: M3536M6 PMS Number: 861465 Part Number: M3536MC
  • Product Description
    Defibrillator/Monitor
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturers representative
    Al-Faisaliah Medical System, Riyadh, (01) 2119948
  • Source
    SFDA