According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by GE Healthcare.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 046 02 18 000
  • Event Initiated Date
    2018-02-13
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    Specifically, in-use experience and limited testing has revealed that the previously communicated disinfection methods may not always be effective when used in connection with procedures that have long tee probe intubation times, such as cardiac surgeries, and/or with tee probes with visible signs of wear or damage in areas to be disinfected, both of which may increase the bioburden on the probe. under these conditions, ge recommends the use of a sterile sheath covering the tee probe in addition to disinfection using the soak based disinfecting method or the tristel trio wipes system as described below.(see attached fsn).

Device

  • Model / Serial
    GE TEE probes used with Tristel Trio Wipes System The following TEE probe models: 6VT-D, 6Tc, 6Tc-RS, 6T, 6T-RS, 9T and 9T-RS. (See Attached FSN for full details)
  • Product Description
    Transducer assembly
  • Manufacturer

Manufacturer