    Recall / Field Safety Notice
  • Event ID
    mdprc 062 07 17 000
  • Event Initiated Date
    2017-07-17
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    There is the potential for damage to the balloon or catheter due to excessive force being applied, and this may lead to balloon inflation or deflation challenges. based on the root cause analysis performed, the failure mode is triggered by the presence of uncured hydrophilic coating in the tip lumen which may cause the stylette to stick to the distal end of the catheter tip.

Device

  • Model / Serial
    Lot#: 212553130 to 213283557 Model numbers : please check attached file
  • Product Description
    Angioplasty catheter, balloon dilatation, coronary, basic
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Manufacturers representative
    Medtronic Saudi Arabia