translation missing: en.activerecord.attributes.event.titles.recall_field_safety_notice

According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by Pro Med Instruments GmbH.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 013 11 17 000
  • Event Initiated Date
    2017-11-06
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    Up to this date, pro med instruments became aware of two breakages of the pin tips, which occurred during two identical surgery procedures, both with the same physician. the breakage occurred at the very distal end of conical part of the skull pin tip. these two breakages did not cause any injury for the patient and/or any third party.

Device

  • Model / Serial
    Item No. 3006-00 LOT 17041 and LOT 17072
  • Product Description
    Disposable Skull Pins, Stainless Steel, Adult
  • Manufacturer

Manufacturer