According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by Philips Medical Systems.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 145 03 14 000
  • Event Initiated Date
    2014-03-31
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    The footboard of patient support for stitching can be fixed in upper position. when the fix clamp (hook) is not correctly adjusted into its correct position, the footboard is not securely held. if it falls down, it may hurt the operator's or patient's foot (e.G. when moving the stand or when the operator fails to fix it correctly upon moving the board up). following factors are needed that the situation occurs: -hook not correctly adjusted -operator does not recognize that the hook does not click into place -patient support is moved (wheeled around) -hook does not hold -foot in “hazard area”.

Device

  • Model / Serial
    All Patient support for stitching are affected
  • Product Description
    X-ray System.
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturers representative
    Al-Faisaliah Medical System, Riyadh, (01) 2119948