translation missing: en.activerecord.attributes.event.titles.recall_field_safety_notice

According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by Accuray Inc..

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 015 10 18 000
  • Event Initiated Date
    2018-10-07
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    Accuray incorporated (accuray) has been informed of two issues related to components of the robot contained within the cyberknife® m6™ system. the issues involve the wrist assembly which is the final linkage between the robot arm and the linear accelerator. one issue relates to the fastening process for the wrist assembly which may not have been implemented properly for specific units. the second involves incorrect material being used in a small number of wrist castings. these issues may lead to premature wrist failure, resulting in a linear accelerator fall.

Device

  • Model / Serial
    Only Accuray CyberKnife M6 Systems may potentially be affected by these issues.
  • Product Description
    Stereotactic Systems, Image-Guided, Radiosurgical, Linear Accelerator
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturers representative
    Gulf Medical Co.
  • Source