translation missing: en.activerecord.attributes.event.titles.recall_field_safety_notice

According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by Medtronic.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 024 05 18 001
  • Event Initiated Date
    2018-05-30
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    This voluntary recall is being conducted due to the potential for a device to be missing the sled component. the sled component is responsible for staple deployment. without the sled, no staples would be deployed. use of a product with this issue may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents. medtronic has received four reports of injury related to this issue.

Device

  • Model / Serial
    Multiple Items and Lots are provided in the attached FSN. ........................ NCMDR update codes: Covidien Endo GIA™ Articulating Reloads. Item code: EGIA45AMT, EGIA60AMT, EGIA60AXT, EGIA60AVM, SIG60AXT, SIG30AVM, EGIA45AV, EGIA45AXT, EGIA45AVM, EGIA45CTAVM, SIG30AMT, SIG30CTAVM, SIG45CTAVM, SIG30CTAV & SIG45AXT See attachment for details*
  • Product Description
    Surgical staple, non-biodegradable
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Manufacturers representative
    Medtronic Saudi Arabia
  • Source
    SFDA