According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by Medtronic Inc.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 011 12 13 000
  • Event Initiated Date
    2013-12-02
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    Medtronic has determined that the following situations can lead to nose cone separations: using the corevalve accutrak dcs to retrieve (remove) a partially deployed valve. removing the corevalve accutrak dcs from the patient, after valve deployment, without fully closing the corevalve accutrak dcs capsule. continuing to pull on the corevalve accutrak dcs if increased resistance is felt at the vessel introducer, the introducer’s hemostatic valve, or other structure during corevalve accutrak dcs retrieval.

Device

  • Model / Serial
    Models DCS-C4-18Fr and DCS-C4-18Fr-23
  • Product Description
    Intended to deliver CoreValve Transcatheter Aortic Valve, which is designed to treat severe aortic valve stenosis without open-heart surgery or surgical removal of the native valve.
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Manufacturers representative
    Gulf Medical Co., (012) 6512828