translation missing: en.activerecord.attributes.event.titles.recall_field_safety_notice

According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by Roche Diagnostics.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 001 07 18 000
  • Event Initiated Date
    2018-07-01
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    Roche has confirmed that sample material may potentially come in contact with the pipetting nozzle during operation. the possible presence of biological material on the nozzle has the potential to cause contamination resulting in potential false positive results, depending on the sensitivity of the analytical technology. however, no complaints of false positives have been received from customers. cobas p 612 (63x) (lcp1) systems that use disposable filter tips are not affected by a potential contamination as the filter tips create a physical barrier preventing liquid exposure to the nozzle.

Device

  • Model / Serial
    GMMI / Part No: 07563116001 SW Version: All versions from 2.0 and higher
  • Product Description
    IVD, Specimen processing instrument IVD, automated
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturers representative
    FAROUK, MAAMOUN TAMER & COMPANY
  • Source
    SFDA