According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by Abbott.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 083 11 17 000
  • Event Initiated Date
    2017-11-23
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    The cell-dyn ruby allows the display of parameter results in usa, international system of units (si), modified international system of units (si mod), unit set 1 or unit set 2 per user configuration. abbott has identified the following issue related to unit set selection for celldyn ruby analyzers using system software versions 2.2ml and lower: standard deviation (sd) values for select parameters (refer to table a) in the qc views only (qc-qcid data view and qc-levey jennings) will be incorrectly displayed when unit sets other than the usa unit are selected on the cell-dyn ruby. these units are incorrectly converted from usa units to other unit sets for display purposes. means, percent cv, and limit (upper and lower) information in qc views are correct. in addition, instrument flagging, alerts, and results are correct for all unit sets.

Device

  • Model / Serial
    List Number: 08H67-01; 08H67-03 Serial Numbers: All serial numbers below 70594BG UDI: 00380740017170; 00380740099916
  • Product Description
    CELL-DYN Ruby
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Manufacturers representative
    Medical supplies & Services Co.Ltd Mediserv