translation missing: en.activerecord.attributes.event.titles.recall_field_safety_notice

According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by Radiometer Medical.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 159 09 12 000
  • Event Initiated Date
    2012-09-25
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    The root cause for the carry-over has been identified to be a deficiency in some cup wash units. that is, the steel tube, which empties the cup during the cup wash process, is too long causing it to touch the bottom of the cup. in this situation an un-bound tracer-analyte complex can bind to the tube and carry-over into subsequent tests. the countermeasure, which reduces the length of the steel tube by 0.4mm to ensure that there is a clearance between the tube and the bottom of the cup.

Device

  • Model / Serial
    AQT90 FLEX: 393-838R0143N0001 Cup wash unit: 903-119 (shipped as of October 31)
  • Product Description
    IVD, analyzers
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturers representative
    Salehiya Trading Est., Riyadh, (01) 2628939
  • Source
    SFDA