translation missing: en.activerecord.attributes.event.titles.recall_field_safety_notice

According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by Abbott.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 069 02 18 000
  • Event Initiated Date
    2018-02-15
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    Abbott has identified the following issues with the alinity ci‐series: 1. after a failure of the reagent 2 (r2) pipettor, the software fails to send scheduled tests that were aspirated after the failure to exceptions. 2. results, reported in alternate result units, that are greater than the linearity incorrectly displays the numerical result value in the default units, for assays using the spline or logit‐4 data reduction method. 3. the host connection is disabled after 12 seconds when no order is received, even though the configured host query time‐out has not been exceeded. 4. an unexpected software shutdown can occur when printing the assay parameter report when the thousand/decimal separator number format is set to the period and comma option. patient impact refer to appendix a for details concerning any patient or safety impact related to the issues identified in alinity ci‐series system software version 2.00.

Device

  • Model / Serial
    03R70‐01
  • Product Description
    Integrated Clinical Chemistry & Immunoassay System
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Manufacturers representative
    Medical supplies & Services Co.Ltd Mediserv
  • Source
    SFDA