translation missing: en.activerecord.attributes.event.titles.recall_field_safety_notice

According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by Mobius Imaging, LLC.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 017 10 17 000
  • Event Initiated Date
    2017-10-02
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    Mobius imaging believes, due to an issue with the configuration settings of the tilt-drive motion controller, if a tilt-drive motor has an intermittent connection or broken encoder or sensor wire there is a risk of unintended tilt motion of gantry while airo system is homing the axis at startup or preparing for a scan. since the airo is not currently monitoring for intermittent connections or broken tilt-drive motor wires, if a wire breaks or the signal becomes intermittent the tilt-drive motor would not get the correct feedback and the motor could behave unpredictably.

Device

  • Model / Serial
    AIRO Mobile CT Scanner (model# MobiCT-32) All Airo systems
  • Product Description
    CT Scanner
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturers representative
    Al-Jeel Medical & Trading Co. LTD
  • Source
    SFDA