translation missing: en.activerecord.attributes.event.titles.recall_field_safety_notice

According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by B Braun Medical Inc.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 004 11 15 001
  • Event Initiated Date
    2015-11-08
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: BfArM
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    During continuous post-market surveillance we became aware of a disconnection of the female luer-lock connector from the venous side of the single-needle adapter leading to blood loss for the patient.

Device

  • Model / Serial
    Art. No. 7211034 Batch # 15C20980 …………………………… BfArM update code : 7211032 : 15C16980 7211033 : 15C23980 7211035 : 15C16980 7210868 : 15D27880
  • Product Description
    Dialysis, A/V Set
  • Manufacturer

Manufacturer