  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 001 06 16 000
  • Event Initiated Date
    2016-06-01
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    The current versions of mr750 dv24, dv25, and dv25.1 software could allow for 3.0t 6 channel flex coil overheating when the device is used in mode 2 setup. this could lead to a serious patient thermal injury.

Device

  • Model / Serial
    Discovery 3.0T MR750 systems running Application software versions DV24 R01, DV25 R02, or DV25.1 M3, when used with the 3.0T 6 Channel Flex Coil.
  • Product Description
    Scanning Systems, Magnetic Resonance Imaging, Extremity
  • Manufacturer

Manufacturer