Field Safety Notices about avance, avance cs2 and aisys anesthetic machines

According to Office for Registration of Medicinal Products, Medical Devices and Biocidal Products, this field safety notices involved a device in Poland that was produced by GE Healthcare.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Field Safety Notice
  • Event ID
    1898
  • Date
    2013-10-31
  • Event Country
  • Event Source
    ORMPMDBP
  • Event Source URL
  • Notes / Alerts
    Polish data is current through November 2018. All of the data comes from the Office for Registration of Medicinal Products, Medical Devices and Biocidal Products, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Poland.
  • Extra notes in the data
  • Action
    GE Healthcare safety note regarding the possibility of providing an excessive tidal volume in Pressure Control Ventilation - Volume Guarantee (PCV-VG) in Avance, Avance CS2 and Aisys anesthetic machines

Device

Manufacturer