  • Type of Event
    Recall
  • Event ID
    18653
  • Event Initiated Date
    2015-06-18
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Teleflex Medical New Zealand, 12 Victoria Street, Private Bag 31346, Lower Hutt, Wellington 5040
  • Reason
    Verathon, manufacturer of glidescope video laryngoscopes, is revising the operations and maintenance manuals (omm) for certain glidescope gvl and avl reusable laryngoscope blades. the new omms specifically state compatibility limitations of the reusable video laryngoscope validated through simulated use cycles, using certain cleaning and disinfection agents. also, for the purpose of cleaning/disinfecting reusable blades, the following chemicals have been removed from the omms:, · bleach (sodium hypochlorite), · isopropyl alcohol solution, · liquid hydrogen peroxide (7.5%).
  • Action
    Instructions for use to be updated

Device

  • Model / Serial
    Model: Models: GVL3, GVL4, GVL5, AVL2, AVL3, AVL4, AVL5, Affected: Selected serial number ranges
  • Manufacturer