Recall of Varian RPM Respiratory Gating System

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Varian Medical Systems Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    14570
  • Event Initiated Date
    2013-04-29
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Varian Medical Systems Australasia Pty Ltd (NZ), C/- BDO Auckland - Christine Koefoed, Level 4, BDO Centre, 4 Graham Street, Auckland 1140
  • Reason
    Manufacturer has identified an anomaly with the rpm respiratory gating system. while operating with the 6-dot marker block, this version of the rpm gating may produce an inverted respiratory waveform in amplitude and phase-gating sessions., the error may be reproduced if the rpm system is induced to frequently lose and re-acquire tracking of the 6-door marker block during a calibration sequence, a reference session or a single session after tracking has begun.
  • Action
    Software to be upgraded

Device

  • Model / Serial
    Model: H50, Affected: 1.7.5 with 3D option SN's: H501056, H501080, H501944, H502289, H502300, H502301, H503567, H503596, H503935, H504201 and H504724.
  • Manufacturer

Manufacturer