Recall of Uninterruptible Power Supply (UPS) for Philips CT / Advanced Molecular Imaging Systems

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Philips Medical Systems (USA).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    18335
  • Event Initiated Date
    2015-04-09
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • Reason
    Philips has become aware of ups device failures that resulted in leaking of electrolyte (sulfuric acid), overheating and/or emission of fumes. analysis has identified that these failures are most likely due to batteries within the ups devices being held in service beyond their usable service life - typically three to five years.
  • Action
    Manufacturer to issue advice regarding use

Device

  • Model / Serial
    Model: BrightView X / XCT, Cardio MD, Forte, Gemini TF/ GXl (6 slice) / TF Astonish / 16 Power / GXL / TF 64 w/ TOF Performance, Precedence, Brilliance CT 16 Slice Air / 64 Channel / Big Bore / iCT / iCT - BETA TEST SYSTEM / iCT SP, Ingenuity CT / Core / Core128, Affected: All units