Recall of Treatment planning system RayStation

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by RaySearch Laboratories AB.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    20365
  • Event Initiated Date
    2016-07-07
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: CARSL Consulting, 50B Puketapu Road, Taradale, Napier 4112
  • Reason
    An issue has been identified where the biological clinical goals for an adapted plan based on another planning ct (rather than the original planning ct) are evaluated using incorrect dose values and therefore display incorrect values.
  • Action
    Manufacturer to issue advice regarding use

Device

  • Model / Serial
    Model: Raystation, Affected: , Software version: 3.0, 3.5, 4.0, 4.7, and 5.
  • Manufacturer

Manufacturer