Recall of TCVS Terumo/Sarns Disposable Centrifugal Pump

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Terumo Cardiovascular Systems Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    16613
  • Event Initiated Date
    2014-05-01
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Terumo Australia Pty Ltd (NZ), c/- Healthcare Logistics, 56 Carrington Road, Mount Albert, Auckland 1025
  • Reason
    Manufacturer has received complaints of the sarns disposable centrifugal punps leaking due to a crack in the molded separator. in all but one report, the leaks were discovered during priming of the cardiopulmonary (cpb) circuit and prior to initiating cpb. in these instances the centrifugal pump was successfully changed out with no delays in the surgical procedures., in one report the failure occurred during cpb and the pump was not replaced.
  • Action
    Product to be returned to supplier