Recall of Synthes_Fixation system implantation kit, internal [44759]

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Synthes GmbH.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    20956
  • Event Initiated Date
    2016-12-16
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Johnson & Johnson (NZ) Limited, 13a Gabador Place, Mt Wellington, AUCKLAND
  • Reason
    The manufacturer received product complaints for breakage in the pfna blade impactor, where the handle detached from the shaft of the instrument.
  • Action
    Product to be exchanged

Device

  • Model / Serial
    Model: , Affected: Impactor for PFNA Blade (Part: 03.010.410 All Lots
  • Manufacturer

Manufacturer