Recall of Synthes DePuy Intramedullary (SP2 IM) Rod 400mm Instrument

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by DePuy Orthopaedic Technology Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    19195
  • Event Initiated Date
    2015-09-25
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Johnson & Johnson (NZ) Limited, 13a Gabador Place, Mt Wellington, AUCKLAND
  • Reason
    It has been determined that there is potential for sp2 im rod 400mm instruments (96-6120) manufactured with 455 stainless steel (ss) or 17-4 ss to fracture and leave a portion of the rods in patients.
  • Action
    Product to be returned to supplier

Device

  • Model / Serial
    Model: 96-6120, Affected: Dates of manufacture: 4-Apr-1995 to 22-Jun-2001, and 27-Aug-2008 to 2-Oct-2012
  • Manufacturer

Manufacturer