Recall of Siemens Sysmex CS-2000i, CS-2100i, CS-2500 and CS-5100 Systems

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Siemens Healthcare Diagnostics Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    21041
  • Event Initiated Date
    2018-07-10
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
  • Reason
    The manufacturer has confirmed that under very rare circumstances a false negative result may be reportable. the issue may only occur under the following conditions:, 1. the initial result is falsely flagged with an antigen excess error that triggers an automatic reanalysis in a 1/19 dilution., 2. the raw signal that is generated from the reanalysis is below the limit of blank (lob) of the method., follow-up communication, the manufacturer has issued a new protocol disc version with new and enhanced assay applications for sysmex cs-2000i/cs-2100i systems. this also includes a new protocol disc version with new and enhanced assay applications for sysmex cs-2500 systems and on sysmex cs-5100 systems.
  • Action
    Manufacturer to issue advice regarding use

Device

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source