Recall of Siemens syngo RT Oncologist software

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Siemens AG.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    19482
  • Event Initiated Date
    2015-11-25
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
  • Reason
    The issue occurs if one or more treatment plans with at least 2 isocenters based on one planning ct are used, and when a cone beam ct has been acquired for each isocenter., the system supports the workflow correctly until the user performs a manual change to another isocenter. at this point the visualization of the overlying images in adaptive targeting might be incorrect. if in this case the user performs a manual registration of the overlying images, this may result in an incorrectly calculated offset in adaptive targeting., applying this offset to the actual patient position can lead to delivering dose to the wrong location and therefore to patient mistreatment.
  • Action
    Software to be upgraded

Device

  • Model / Serial
    Model: 7356335, Affected: , Software version: 4.3.138, 4.3.SP1, 4.3.1_AR1, 4.3.1_MR2