Recall of Siemens Immulite 2000 and Immulite 2000 XPi, Reagent Carousel Temperature Sensor

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Siemens Healthcare Diagnostics Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    18449
  • Event Initiated Date
    2015-04-29
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
  • Reason
    Siemens has identified that affected systems have been manufactured with a defective reagent carousel temperature sensor pcb or may have had a part replacement with a defective pcb assembly., a particular set of the temperature sensor boards were built incorrectly. the incorrectly assembled temperature sensor boards may cause icing on the reagent carousel casting. if there is icing within the reagent carousel, there is a potential for the reagent carousel to encounter a mechanical jam which will stop the system from processing samples and will generate errors.
  • Action
    Product to be modified