Recall of Siemens Flex Large 4 Coil for MAGNETOM Skyra, Verio, Spectra and Biograph mMR

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Siemens AG.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    13497
  • Event Initiated Date
    2012-09-14
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
  • Reason
    Heating may occur if each of the following conditions applies at the same time:, - coil positioned off-center in head-foot direction by more than +/- 5 cm,, - coil at side position, - application of the high rf power over several minutes, in that position , the surface temperature may well exceed temperatures defined by regulatory standards and may cause local burns to the patient.
  • Action
    Instructions for use to be updated

Device

  • Model / Serial
    Model: Part No. 8625761, Affected: All units