Recall of Siemens Dimesnion Vista System

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Siemens Healthcare Diagnostics Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    19942
  • Event Initiated Date
    2016-03-24
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Siemens Healthcare (NZ) Ltd, 55 Hugo Johnston Drive, Penrose, Auckland
  • Reason
    Testing with a new lot of antibody showed on-board stability claim of 20 hours for dimension vista methotrexate application was not met. lots g2 and h1 (current antibody) showed similar results. further testing on lot g2 showed 4 hours on-board stability.( description of issue – siemens healthcare diagnostics has confirmed that the syva® emit® methotrexate application for the dimension vista® system does not meet the on-board stability claim of 20 hours).
  • Action
    Software to be upgraded