Recall of Siemens Centaur Tnl Ultra

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Siemens Healthcare Diagnostics Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    18574
  • Event Initiated Date
    2015-05-22
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
  • Reason
    Siemens healthcare diagnostics has observed a system-to-system bias between the tni-ultra™ assay on the advia centaur® cp and advia centaur/advia centaur xp/advia centaur xpt systems with the advia centaur cp generating lower results than the advia centaur/advia centaur xp/advia centaur xpt system. siemens is actively pursuing the cause of this issue., this observation impacts customers who use the tni-ultra assay on both the advia centaur cp and advia centaur/advia centaur xp/advia centaur xpt systems interchangeably when interpreting serial testing in patient samples. this communication does not impact customers who use tni-ultra results solely from either the advia centaur/advia centaur xp/advia centaur xpt system or the advia centaur cp system.
  • Action
    Manufacturer to issue advice regarding use

Device

  • Model / Serial
    Model: SMN: 10317708 (100 test kit), 10317709 (500 test kit), Affected: Kit Lots ending in 088, 089, 090, 091, 093 and 094 and all future lots until the issue is resolved and a follow-up communication is issued
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    NZMMDSA