Recall of Siemens Capillary Caps for RapidLyte Mulitcap and Multicap-S Blood Collection Capillaries

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Siemens Healthcare Diagnostics Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    18045
  • Event Initiated Date
    2015-01-30
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
  • Reason
    A material change made by the supplier has resulted in capillary caps that are more difficult to apply to the capillaries. attempting to apply these capillary caps to the capillaries could result in spillage of the sample or possible breakage of the glass capillary.
  • Action
    Manufacturer to issue advice regarding use

Device

  • Model / Serial
    Model: Part Numbers: 478601 (SMN 10311053), 478605 (SMN 10311054), 478527 (SMN 10328655), Affected: All lots
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    NZMMDSA