Recall of Siemens ADVIA Chemistry Calibrator

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Siemens Healthcare Diagnostics Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    20749
  • Event Initiated Date
    2016-10-19
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
  • Reason
    The dbil_2 assigned calibrator values (acv) may cause, on average, a positive proportional bias of approximately 41% on patient samples compared to the reference method for direct bilirubin measurement. due to the proportional bias, the differences become greater as the concentration of dbil_2 increases. quality control (qc) material is also affected by this issue. depending on the quality control concentrations and ranges used by the laboratory, quality control may not detect this issue.
  • Action
    Manufacturer to issue advice regarding use