Recall of Siemens Advia Centaur Systems Progesterone Kit

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Siemens Healthcare Diagnostics Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    18048
  • Event Initiated Date
    2015-01-30
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
  • Reason
    Siemens healthcare diagnostics has confirmed a high bias on results greater than 30 ng/ml (95.4 nmol/l) when compared to the master curve when using affected advia centaur® progesterone kit lots. the high bias may cause neat serum samples with results greater than 60 ng/ml (190.8 nmol/l) to recover within the assay range upon dilution (dilution factor applied). on the advia centaur systems, for samples greater than 30 ng/ml (95.4 nmol/l), customers may observe an average dilution recovery between 48.3% and 105.0%.
  • Action
    Manufacturer to issue advice regarding use