Recall of Siemens ADVIA Centaur FSH Assay, Calibrator B

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Siemens Healthcare Diagnostics Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    13310
  • Event Initiated Date
    2012-08-17
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
  • Reason
    The manufacturer advises that the analytical sensitivity values greater than 0.3 mlu/ml described in the instructions for use (ifu) when evaluating multiple lots of the advia centaur fsh assay and calibrator b on the advia centaur and advia centaur xp system.
  • Action
    Product to be destroyed

Device

  • Model / Serial
    Model: 00649625, 00652707, 00654254, 00650577, 00654130, Affected: Lots ending in; 29 (expiry date 2012-12-02); 30 (expiry date 2012-12-02); 40 (expiry date 2013-04-09)
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    NZMMDSA