Recall of RaySearch RayStation Treatment Planning System

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by RaySearch Laboratories AB.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    21046
  • Event Initiated Date
    2017-01-25
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: CARSL Consulting, 50B Puketapu Road, Taradale, Napier 4112
  • Reason
    Investigation showed that in a certain, non-standard, remoting installation, accelerated computation on gpu may be wrong for a very large computation problem. this can affect deformable registration and dose computation., the error affectes optimization and there is a risk that the error drives the optimizer into creating a plan that apears acceptable although there is unacceptable dose in some region.
  • Action
    Manufacturer to issue advice regarding use

Device

Manufacturer