Recall of Radiometer ABL90 Flex analyser

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Radiometer Medical ApS.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    15623
  • Event Initiated Date
    2013-11-08
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Radiometer Pacific, Unit 3, 33 Spartan Road, Takanini 2105
  • Reason
    Radiometer has recently become aware of a potential issue of negative bias on sodium results if the abl90 analyser is not on a horizontal table. the error has only been reported when the analyzer has been installed on a roll stand and tilted (i.E. the support plate is not horizontal) and the analyser has been idle for more than 30 minutes.
  • Action
    Software to be upgraded

Device

Manufacturer