Recall of Q Core Sapphire Infusion Pump

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Q Core Medical Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    18872
  • Event Initiated Date
    2015-07-21
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Hospira NZ Ltd, 58 Richard Pearse Drive, Airport Oaks, Mangere, Auckland 2022
  • Reason
    The manufacturer has received complaints from customers that when using the epidural mode with patient bolus, the amount of medication delivered may be greater than programmed. q core medical is adding to the user manual the parameters of the reference catheter to allow precise correlation between the catheter dimensions, the back pressure and the deviation in flow that the catheter may cause.
  • Action
    Instructions for use to be updated