Recall of Physio Control Lifepak 1000 Defibrillator Automated

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Physio-Control Corporation.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    17001
  • Event Initiated Date
    2014-07-28
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Medxus Ltd, Unit 4, 100 Fitgerald Avenue, Christchurch 8011
  • Reason
    Physio-control has become aware of incidents where customers have attempted to use their lifepak 1000 defibrillator and the device has shut down unexpectedly due to a very low battery. a defibrillator in this scenario has the potential to fail to deliver a shock, with the potential result that therapy is not delivered and a patient is not resuscitated., physio-control has determined this to be as a result of batteries not being replaced when they have reached a low or very low state of charge as indicated in the readiness display on the device.
  • Action
    Instructions for use to be updated