Recall of Philips IntelliVue Patient Monitors

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Philips Medizinsysteme Boblingen GmbH.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    20465
  • Event Initiated Date
    2016-08-04
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • Reason
    It has recently been determined that when certain patient monitors (mx400, mx430, mx450, mx500, mx550) have been powered on continuously for at least three months, the waveforms displayed on the monitor may be outdated and therefore fail to reflect the patient's current condition. should this occur, however, the monitor's alarms and numeric displays of vital signs will continue to function as specified and accurately reflect current patient data.
  • Action
    Software to be upgraded