Recall of Philips IntelliVue Patient Monitors

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Philips Medizinsysteme Boblingen GmbH.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    17594
  • Event Initiated Date
    2014-10-30
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • Reason
    In time-synchronized automatic/sequence mode, the non-invasive blood pressure (nbp) automatic measurement series is stopped inadvertently if the repetition interval is increased manually by the user (e.G. from 5 to 15 minutes) in between two consecutive measurements. if the nbp measurement is stopped, and a patient's blood pressure is not being monitored, a delay in treatment could occur.
  • Action
    Software to be upgraded