Recall of Philips Intellivue Information Center (PIIC iX) Central System

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Philips Medical Systems (USA).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    22413
  • Event Initiated Date
    2018-01-08
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • Reason
    Software issue. once a surveillance station is restarted on january 1, 2018 or later, the station will be unable to perform patient discharge and transfer operations. any subsequent attempt to perform these operations will cause the station to restart, resulting in a short period of loss of monitoring at the surveillance station during such restart.
  • Action
    Software to be upgraded

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    NZMMDSA