Recall of Philips HeartStart MRx Monitor/Defibrillator

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Philips Medical Systems (USA).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    17839
  • Event Initiated Date
    2014-12-11
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • Reason
    The mrx could contain incorrect internal software settings, causing the following two issues:, 1. the device will perform the weekly automated tests hourly, which could cause the therapy capacitors to degrade sooner than intended., 2. while connected to ac or dc power and with no battery installed or the battery installed has a charge level of less than 10%, the ready for use (rfu) indicator will not provide the expected low battery indication (flashing red x with audible chirp). instead, the rfu will show a flashing black hourglass, indicating that sufficient battery power is available for device operation.
  • Action
    Software to be upgraded