Recall of Philips Heartstart MRx Monitor/Defibrillator

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Philips Medical Systems (USA).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    14847
  • Event Initiated Date
    2013-06-18
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • Reason
    The philips heartstart mrx monitor/defibrillator can deliver a non- synchronised shock in the synchronised cardioversion mode. this can occur if the user rotates the therapy knob while simultaneously pressing the sync button, then charges the mrx and presses the shock button., the mrx will still display r wave markers as if normally operating in synchronised cardioversion mode. the "sync" label will still be seen on the display. when the mrx is charged and the shock button pressed, the shock will be immediately delivered and may not be synchronised to the r wave., in addition, the label affixed to the top of the mrx device is not consistent with the mrx instructions for use on synchronised cardioversion.
  • Action
    Software to be upgraded

Device

  • Model / Serial
    Model: Models: M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6, Affected: Serial numbers within the range US00100100 to US00567299