Recall of Philips Heartstart MRx Monitor/Defib, lithium ion batteries

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Philips Medical Systems (USA).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    16286
  • Event Initiated Date
    2014-02-24
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Laerdal (New Zealand) Ltd, Unit O, Cain Commerical Centre, 20 Cain Road, Penrose
  • Reason
    The philips heartstart mrx monitor/defibrillator, when operating on battery power only, may experience an unexpected shutdown if exposed to elevated levels of electromagnetic interference from rf energy sources., examples of equipment that can cause electromagnetic interference are medical devices, cellular products, information technology equipment and radio/television transmissions.
  • Action
    Product to be exchanged

Device

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    NZMMDSA