Recall of Philips Heartstart MRx Monitor/Defib

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Philips Medical Systems (USA).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    16285
  • Event Initiated Date
    2014-02-24
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • Reason
    When used in hospital transport and pre-hospital (ems) environments, the connection between the pads/paddles therapy cable (including pads cpr therapy cable, external paddles cable, and internal paddles cable/adapter) and the mrx therapy connection port could experience high levels of stress causing accelerated wear. wear of the connection could prevent the device from sensing that the pads/paddles therapy cable is connected. this wear also could cause the mrx to inappropriately identify the pads therapy cable, external paddles, or internal paddles., pads/paddles therapy cable and therapy connection port wear could pose a risk for patients and/or caregivers.
  • Action
    Product to be modified

Device

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    NZMMDSA