Recall of Philips Heartstart MRx Monitor/Defib

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Philips Medical Systems (USA).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    16283
  • Event Initiated Date
    2014-02-24
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Laerdal (New Zealand) Ltd, Unit O, Cain Commerical Centre, 20 Cain Road, Penrose
  • Reason
    A component of the mrx processor board may be susceptible to damage from electrostatic discharge (esd). damage to these components can disrupt ecg and spo2 functionality leading to an inability to perform:, " demand mode pacing, " 12 lead ecg acquisition and analysis, " ecg monitoring via leads, " pulse oximetry (spo2), the presence of this condition will be displayed on the ready for use (rfu) indicator as a solid red x with periodic audio chirp., defibrillation, fixed mode pacing, ecg monitoring via pads/paddles, and other monitoring functions are not impacted by this issue.