Recall of Ortho Reagent Red Blood cell Products

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by manufacturer #1479.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    20924
  • Event Initiated Date
    2016-12-08
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Ortho-Clinical Diagnostics, c/o Quigg Partners, Level 7, 36 Brandon Street, Wellington 6011
  • Reason
    Manufacturer has received an increase in customer complaints and has confirmed the intermittent presence of marked hemolysis in ortho reagent red blood cell (rrbc) products as well as quality control products containing red blood cells., an extensive investigation identified microbial contamination as the root cause for the marked hemolysis.
  • Action
    Manufacturer to issue advice regarding use

Device

  • Model / Serial
    Model: 719102, 6902317, 6902315, 6902316, 719402, 719522, 719201, 6902096, 719502, 719810, 6902318, 719210, 6902319, 6901865, 719510, 707910, 7192042, 719602, 719100, 7192041, 6908163, 719610, 707950, 7191041, 719221, 719202, 707930, Affected: Multiple lot numbers
  • Manufacturer

Manufacturer