Recall of Ortho-Clinical VITROS Chemistry Product TIBC Kit

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Ortho-Clinical Diagnostics Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    14255
  • Event Initiated Date
    2013-02-22
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Ortho-Clinical Diagnostics c/- Johnson & Johnson, 13A Gabador Place, Mt Wellington, AUCKLAND 1060
  • Reason
    Manufacturer has become aware of discrepancies between the correct measuring ranges provided in the vitros chemistry products instructions for use and those entered on the vitros system for calibration diskettes (cal disks) and assay data disks (add) for these products.
  • Action
    Manufacturer to issue advice regarding use

Device

Manufacturer