Recall of NucliSENS easyMag Magnetic Silica Mag SIL

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Biomerieux SA.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    20575
  • Event Initiated Date
    2016-08-30
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Medi'Ray New Zealand Ltd, 53-55 Paul Matthews Road, Albany, Auckland 0632
  • Reason
    Following customer complaints about amplification performance issues with the home brew polymerase chain reaction (pcr) and real time pcr assays. the investigation concluded that downstream applications were impacted when the extraction volume was higher than 400 ìl and for double stranded nucleic acids target., double stranded nucleic acid applications with small (< 40kbp) and medium genome sizes (< to 1200 kbp) i.E dna viruses are more impacted than higher human genomic dna and bacterial applications (> to 1200kbp.
  • Action
    Manufacturer to issue advice regarding use